Factory Calibration

Recommended Service Frequency: No regular factory service required.

What's included with Smart Chamber factory service?

A factory recalibration includes the following tests and maintenance procedures:

  • Test and verify Wi-Fi, USB, and Ethernet connections can communicate with a computer.
  • Test USB and Ethernet connections to instruments.
  • Test power out port.
  • Test soil probe port.
  • Test thermistor port.
  • Test circulation flow.
  • Inspect bellows and check bellows pressure.
  • Check GPS receiver.
  • Take a sample measurement and verify data storage.
  • Update device firmware.
  • Check soil collar gasket and replace if needed.
  • Check foam soil collar gasket and replace if needed.
  • Check chamber gasket and replace if needed.
  • Test battery.
  • Replace air filter.
  • Other repairs and maintenance as needed. Customer approval is required for additional repairs.

To request factory service or recalibration for an instrument:

Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.